Product Description

The SVM™ series provides customers more parameters as compared to any other kinematic viscometer on the market. There are 3 different models: SVM™ 2001 is a fast and easy-to-use viscometer for the purpose of measuring kinematic viscosity at any temp. between 15 deg C and 100 deg C, whereas SVM™ 3001 gives measurements between -60 deg C and +135 deg C. An individual measuring cycle on a small specimen volume produces kinematic viscosity, dynamic viscosity, density, Viscosity Index and so on. SVM™ 4001 is a double-cell device especially designed for simultaneous measurements of density and viscosity at 02 different temps. It mechanically provides the Viscosity Index from only a small volume  of sample. Various accessories, such as a hot filling attachment or an automated samplers, and modularity options are accessible for the series.

Key Features

Multiple parameter measurement

SVM™ provides our customers with an exceptional solution that is unique on the market: The possibility to not only measure viscosity, but multiple parameters from only one syringe. From kinematic and dynamic viscosity to density measurements and more – the possibilities are vast. Each parameter measured by SVM™ 3001 or SVM™ 4001 complies with the following industry standards:

  • Kinematic viscosity (ASTM D7042, EN 16896, DIN 51659-2)
  • Density (EN ISO 12185, ASTM D4052, IP 365)
  • Dynamic viscosity (ASTM D7042)
  • Viscosity Index (VI) (ISO 2909, ASTM D2270)
  • API grades (ISO 91, API 2540, ASTM D1250, IP 200)
  • Saybolt viscosity (ASTM D2161)

Unbeatable ease of operation

The SVM™ series stands out for its unbeatable ease of operation, from the initial start-up to the maintenance of the instruments. The devices come factory-adjusted, which means they are ready for immediate use. The measurement is started by simply injecting the sample from a syringe. The SVM™ instruments guarantee safe and easy handling during the measurement procedure, without leaks or breakage. In addition, easy cleaning with a wash bottle simplifies the work.

Low sample and solvent volume

An outstanding feature of the SVM™ series is the small amount of sample and solvent volume required to obtain precise measurement results. This not only simplifies the measuring procedure, but also lowers costs the minimum sample volume is only 1.5 mL and the amount of solvent needed for cleaning is low. Disposal costs can therefore be considerably reduced. Therefore, these SVM™ instruments not only impress with their performance, but also with their sustainability.

Wide temperature range

The instruments of our SVM™ series provide a wide temperature range for both viscosity and density. SVM™ 3001 comes with a range from -60 °C to +135 °C for measuring jet fuel and diesel fuel as well as lubricating oil and wax – with only one integrated cell. Temperatures down to -20 °C can be reached with built-in air cooling (without external cooling), whereas temperatures as low as -60 °C are reached by using external cooling with water/glycol mixtures; no flammable liquids are required. SVM™ 4001 is designed to determine viscosity and density simultaneously at any two temperatures between 15 °C and 100 °C.

One measuring cell – all samples

The SVM™ instruments have one measuring cell for the entire viscosity, density, and temperature range and the sample is measured in one go. This means that samples of different viscosity, such as jet fuels, lubricating oils, heavy fuels and waxes, can all be tested with one integrated measuring cell, which is especially relevant for testing laboratories. In addition, SVM™ 3001 enables you to perform temperature scans and gain knowledge of the sample’s behavior such as low temperature fluidity. Connecting the SVM™ with the 71-position sample changer means you are free to perform other important tasks.

Technical Specifications


SVM™ 2001

SVM™ 3001

SVM™ 4001

Temperature range

+15 °C to +100 °C

-60 °C to +135 °C

+15 °C to +100°C

Viscosity range

0.2 mm²/s to 30 000 mm²/s

Density range

0.6 g/cm³ to 3 g/cm³

Main standards

ASTM D7042, EN 16896

ASTM D7042, EN 16896
ASTM D4052, ISO 12185

ASTM D7042, EN 16896
ASTM D4052, ISO 12185

Supported precision classes

Ultrafast, Fast and Precise

Ultrafast, Fast, Precise and Ultraprecise

Ultrafast, Fast, Precise and Ultraprecise

Sample volume min./typical

1.5 mL / 5 mL

1.5 mL / 5 mL

2.5 mL / 6 mL

Solvent volume min./typical

1.5 mL / 6 mL

1.5 mL / 6 mL

2.5 mL / 10 mL

Maximum sample throughput

30 samples per hour

24 samples per hour

Peltier temperature control

Designed for constant temperature

Designed for fast heating/cooling over a wide range

Designed for simultaneous measurement at any two different temperatures within the range

Optional automation

Single syringe (5 mL or 10 mL) or 45 vials with 35 mL or 71 vials with 12mL

Wetted parts

Inside the instrument: Copper, Titanium, Stainless steel A4, Inconel®

O-rings in contact with sample

Viton® Extreme

Kalrez® Spectrum 0040

Viton® Extreme

Data memory

1000 measurement results

HID (Human Interface Device)

Touchscreen; optional keyboard, mouse and 2D bar code reader

Interfaces

4 x USB (2.0 full speed); 1 x Ethernet (100 Mbit); 1 X CAN bus; 1 x RS-232; 1 x VGA

Power supply

AC 100 V to 240 V; 50 Hz to 60 Hz; 250 VA max.

Ambient conditions

15 °C to 35 °C (59 °F to 95 °F), max. 80% r.h. non-condensing

Net weight/shipping weight

15.9 kg/20.5 kg

17.6 kg/22.2 kg

17.8 kg/22.4 kg

Dimensions (W x D x H)

33 cm x 51 cm x 23.1 cm (13 in x 20 in x 9.1 in)

Compliance

CE mark; EMC directive EN 61326-1; LV directive EN 61010-1; RoHS

Special features

Optional automation

Automatic VI method, API calculations, temperature scans, FillingCheck™. Optional: Countercooling, automation, modularity with Abbemat refractometers

Double measurement cells for simultaneous viscosity and density measurement at any two different temperatures in the range, VI method, API calculations, FillingCheck™. Optional: Automation, modularity with Abbemat refractometers